Skip to main content

Development of a real-time and quantitative thrombus sensor for an extracorporeal centrifugal blood pump by near-infrared light.

Author
Abstract
:

We developed an optical thrombus sensor for a monopivot extracorporeal centrifugal blood pump. In this study, we investigated its quantitative performance for thrombus detection in acute animal experiments of left ventricular assist using the pump on pathogen-free pigs. Optical fibers were set in the driver unit of the pump. The incident light at the near-infrared wavelength of 810 nm was aimed at the pivot bearing, and the resulting scattered light was guided to the optical fibers. The detected signal was analyzed to obtain the thrombus formation level. As a result, real-time and quantitative monitoring of the thrombus surface area on the pivot bearing was achieved with an accuracy of 3.6 ± 2.3 mm2. In addition, the sensing method using the near-infrared light was not influenced by changes in the oxygen saturation and the hematocrit. It is expected that the developed sensor will be useful for optimal anticoagulation management for long-term extracorporeal circulation therapies.

Year of Publication
:
2018
Journal
:
Biomedical optics express
Volume
:
9
Issue
:
1
Number of Pages
:
190-201
Date Published
:
2018
DOI
:
10.1364/BOE.9.000190
Short Title
:
Biomed Opt Express
Download citation