Development of a 125I source for its application in bone densitometry.
Author | |
---|---|
Abstract | :
This paper describes a method for the preparation of (125)I sources for their use in bone densitometer. The process consists of preparation of silver pellets of size ∼2.5 mm (ϕ) × 0.6 mm (l) by 'Cold die compaction technique', palladium coating of the silver pellets, adsorption of (125)I on the palladium coated silver pellets and encapsulation source matrix in a titanium capsule [ 3 mm (ϕ) × 10 mm(l)] using Nd:YAG laser. The texture and morphology of the source matrix were examined by SEM and EDS analyses. Encapsulated sources containing (125)I activity up to ∼1.85 GBq (50 mCi) were prepared. Quality control tests that were necessary before application were performed and gave satisfactory results. |
Year of Publication | :
2012
|
Journal | :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
|
Volume | :
70
|
Issue | :
3
|
Number of Pages | :
470-7
|
ISSN Number | :
0969-8043
|
URL | :
https://linkinghub.elsevier.com/retrieve/pii/S0969-8043(11)00545-8
|
DOI | :
10.1016/j.apradiso.2011.11.010
|
Short Title | :
Appl Radiat Isot
|
Download citation |